(1994) Physics in Medicine and Biology — Vol. 39, n° 4, p. 719-734 (1994)
Files
No attached file found for this publication.
Details
Authors
Walrand, StéphanUCLouvain
Author
van Elmbt, LarryUCLouvain
Author
Pauwels, StanislasUCLouvain
Author
Abstract
A new method for correcting simultaneously the attenuation, scatter and resolution effects in SPECT has been developed for the case of a homogeneous medium. It is based on an effective model of the scatter process. This model depends on only four parameters which are determined experimentally and remain independent of the geometry and of the dimensions of the scatter medium. The method uses the data from the peak events and does not need additional energy windows on the scattered events. An original filter is proposed to remove the noise due to the poor statistics of clinical data. Tests on phantoms varying in size and activity show that the model allows absolute activity determination with an accuracy of a few per cent.
Walrand, S., van Elmbt, L., & Pauwels, S. (1994). Quantitation in SPECT using an effective model of the scattering. Physics in Medicine and Biology, 39(4), 719-734. https://doi.org/10.1088/0031-9155/39/4/005 (Original work published 1994)