[Transfer of an automatic dosage method of noradrenaline in human urine: Utilization of total error as decision critera]

Rozet, E.;Dewe, W.;Chiap, P.;Lecomte, F;Hubert, P.
(2006) 3rd Meeting of the Internationales-de-Toxicologie-Hospitaliere — Location: Liege(Belgium) (24.November.2005)

Files

No attached file found for this publication.

Details

Authors
  • Rozet, E.
    Author
  • Dewe, W.
    Author
  • Chiap, P.
    Author
  • Lecomte, F
    Author
  • Hubert, P.
    Author
Abstract
Two new statistical approaches based on the total error of measurements were applied to the transfer of an automated method for the determination of noradrenatin in human urine by LC-ECD coupled on-line to SPE. They showed that the receiving laboratory was mastering the analytical procedure allowing it to use the method in routine. Furthermore the risk based approach gave guarantee that the risk to have future measurements out of specification in the receiving laboratory was smaller than 5%, therefore being a risk management tool.
Affiliations

Citations

Rozet, E., Dewe, W., Chiap, P., Lecomte, F., & Hubert, P. (2006). [Transfer of an automatic dosage method of noradrenaline in human urine: Utilization of total error as decision critera]. Acta Clinica Belgica (Multilingual Edition), 61, 57-59. https://hdl.handle.net/2078.5/140119 (Original work published 2006)