[Magnetometer for weak fields with resolution of 10 ppm]

Claessens, Ch.;Warnant, H.;Boving, R.;Jones, A.
(1972) Nuclear Instruments and Methods — Vol. 101, n° 2, p. 279-286 (1972)

Files

No attached file found for this publication.

Details

Authors
  • Claessens, Ch.
    Author
  • Warnant, H.
    Author
  • Boving, R.
    Author
  • Jones, A.
    Author
Abstract
This magnetometer has a 10 ppm resolution in measuring fields between 30 and 200 G, performances being independent of room and probe temperature. Measures are achieved by means of nuclear resonance of premagnetised flowing water. The probe contains only 1 cm/sup 3/ of water.
Affiliations

Citations