[Follow-up of TIPS: Assessment of signs of dysfunction at Doppler sonography]
Puttemans, T.;Van Beers, Bernard;Goffette, Pierre;Dardenne, AN.;Pringot, Jacques
(1996) Journal de Radiologie : revue d’imagerie medicale, diagnostique et therapeutique — Vol. 77, n° 12, p. 1201-1206 (1996)
Files
No attached file found for this publication.
Details
Authors
Puttemans, T.
Author
Van Beers, BernardUCLouvain
Author
Goffette, PierreUCLouvain
Author
Dardenne, AN.
Author
Pringot, JacquesUCLouvain
Author
Abstract
Purpose : to assess the signs of TIPS dysfunction at Doppler sonography. Materials and methods : retrospective study of signs observed in 106 TIPS including 31 TIPS with dysfunction (portoauricular pressure gradient >12 mmHg). Results : the signs of TIPS dysfunction were a decrease in the mean velocity in the TIPS (for a velocity <40 cm/sec, 90% sensitivity, 96% specificity), a hepatopetal intrahepatic portal flow (on the right 90% sensitivity, 100% specificity, on the left 95% sensitivity, 92% specificity), a lack of cardiac modulation of the signal in the TIPS (93% sensitivity, 65% specificity), a hepatic vein flow reversal (30% sensitivity, 100% specificity), and, the only direct sign of stenosis, an increase of the mean velocity in the stenosis (42% sensitivity, 95% specificity for a velocity > or = 1 m/sec). Conclusion : the most sensitive Doppler signs of tips dysfunction are the indirect signs of stenosis.
Puttemans, T., Van Beers, B., Goffette, P., Dardenne, AN., & Pringot, J. (1996). [Follow-up of TIPS: Assessment of signs of dysfunction at Doppler sonography]. Journal de Radiologie : revue d’imagerie medicale, diagnostique et therapeutique, 77(12), 1201-1206. https://hdl.handle.net/2078.5/49656 (Original work published 1996)