Technical description of a modified jet ventilation injector for airway laser surgery in neonates and infants: Retrospective analysis of 20 cases

Rosal, Maria;Van Boven, Michel;Schmitz, Sandra;Hamoir, Marc;Veyckemans, Francis
(2016) Journal of Clinical Anesthesia — Vol. 32, n° 32, p. 142-147 (2016)

Files

Hamoir-Airwaylasersurgery.pdf
  • Restricted Access
  • Adobe PDF
  • 661.41 KB

Details

Authors
  • Rosal, MariaUCLouvain
    Author
  • Van Boven, MichelUCLouvain
    Author
  • Author
  • Hamoir, MarcUCLouvain
    Author
  • Veyckemans, FrancisUCLouvain
    Author
Abstract
Introduction The authors modified an adult jet ventilation injector (Hunsaker Mon-Jet Ventilation Tube®) to be able to provide transglottal high-frequency jet ventilation (HFJV) in small children undergoing laryngeal procedures with CO2 laser. Methods and material Retrospective review of the anesthetic records of all children younger than 2 years undergoing transglottal HFJV for CO2 laser laryngeal procedures using this modified adult injector between 2006 and 2013. Results Nine children (5 boys, 4 girls) were identified who underwent a total of 20 procedures. Mean age was 7.4 ± 6.9 months, and mean weight was 6 ± 2.8 kg. No complications were observed with the use of HFJV or this modified injector. Conclusion In experienced hands, this modified injector ensures excellent visibility and field access to the surgeon as well as adequate ventilation during laryngeal laser surgery in infants.
Affiliations

Citations

Rosal, M., Van Boven, M., Schmitz, S., Hamoir, M., & Veyckemans, F. (2016). Technical description of a modified jet ventilation injector for airway laser surgery in neonates and infants: Retrospective analysis of 20 cases. Journal of Clinical Anesthesia, 32(32), 142-147. https://doi.org/10.1016/j.jclinane.2016.01.031 (Original work published 2016)