Low-dose CT allows for accurate proton therapy dose calculation and plan optimization

Elhamiasl, Masouf;Salvo, Koen;Poels, Kenneth;Defraene, Gilles;Nuyts, Johan;et.al.
(2022) Physics in medicine and biology — Vol. 67, n° 19, p. 195015 (2022)

Files

Masoud_Elhamiasl_PMB2022.pdf
  • Open Access
  • Adobe PDF
  • 5.86 MB

Details

Authors
  • Elhamiasl, Masouf
    Author
  • Salvo, Koen
    Author
  • Poels, Kenneth
    Author
  • Defraene, Gilles
    Author
  • Geets, XavierUCLouvain
    Author
  • Author
  • Nuyts, Johan
    Author
Show more
Abstract
(en) OBJECTIVE.Protons offer a more conformal dose delivery compared to photons, yet they are sensitive to anatomical changes over the course of treatment. To minimize range uncertainties due to anatomical variations, a new CT acquisition at every treatment session would be paramount to enable daily dose calculation and subsequent plan adaptation. However, the series of CT scans results in an additional accumulated patient dose. Reducing CT radiation dose and thereby decreasing the potential risk of radiation exposure to patients is desirable, however, lowering the CT dose results in a lower signal-to-noise ratio and therefore in a reduced quality image. We hypothesized that the signal-to-noise ratio provided by conventional CT protocols is higher than needed for proton dose distribution estimation. In this study, we aim to investigate the effect of CT imaging dose reduction on proton therapy dose calculations and plan optimization. APPROACH.To verify our hypothesis, a CT dose reduction simulation tool has been developed and validated to simulate lower-dose CT scans from an existing standard-dose scan. The simulated lower-dose CTs were then used for proton dose calculation and plan optimization and the results were compared with those of the standard-dose scan. The same strategy was adopted to investigate the effect of CT dose reduction on water equivalent thickness (WET) calculation to quantify CT noise accumulation during integration along the beam. MAIN RESULTS.The similarity between the dose distributions acquired from the low-dose and standard-dose CTs was evaluated by the dose-volume histogram and the 3D Gamma analysis. The results on an anthropomorphic head phantom and three patient cases indicate that CT imaging dose reduction up to 90% does not have a significant effect on proton dose calculation and plan optimization. The relative error was employed to evaluate the similarity between WET maps and was found to be less than 1% after reducing the CT imaging dose by 90%. SIGNIFICANCE.The results suggest the possibility of using low-dose CT for proton therapy dose estimation, since the dose distributions acquired from the standard-dose and low-dose CTs are clinically equivalent.
Affiliations

Citations

Elhamiasl, M., Salvo, K., Poels, K., Defraene, G., Lambrecht, M., Geets, X., Sterpin, E., & Nuyts, J. (2022). Low-dose CT allows for accurate proton therapy dose calculation and plan optimization. Physics in medicine and biology, 67(19), 195015. https://doi.org/10.1088/1361-6560/ac8dde (Original work published 2022)