The risk of oversizing with sutureless bioprosthesis in small aortic annulus.

Baert, Jérôme;Astarci, Parla;Noirhomme, Philippe;de Kerchove, Laurent
(2017) The Journal of Thoracic and Cardiovascular Surgery — Vol. 153, n° 2, p. 270-272 (2017)

Files

2017_Baert_J_Thorac_Cardiovasc_Surg.pdf
  • Open Access
  • Adobe PDF
  • 581.73 KB

Details

Authors
  • Baert, JérômeUCLouvain
    Author
  • Astarci, ParlaUCLouvain
    Author
  • Noirhomme, PhilippeUCLouvain
    Author
  • de Kerchove, LaurentUCLouvain
    Author
Abstract
(en) The Perceval S (Liva Nova PLC, London, United Kingdom)sutureless bioprosthesis is a self-expanding bovinepericardial valve. Because of a simple and fast implantationtechnique, this valve allows a short crossclamp time withexcellent hemodynamics and almost no paravalvularleak.1,2The supplied sizers determine the best-suitedprosthesis by measuring the aortic annulus after native valveresection.3We report the case of 4 patients in whomincomplete deployment led to severe dysfunction andexplantation of the Perceval S bioprosthesis [...]