Human somatotropin was assayed by a novel automated nonradioisotopic technique, "particle-counting immunoassay," that requires a 45-min incubation and only 60 microL of 10-fold diluted sample. The principle of the assay is agglutination of antibody-coated latex particle by somatotropin, the reaction being measured by the (instrumented) counting of residual non-agglutinated particles. The dynamic range in serum extends from 0.2 to 40 micrograms/L. The between-assay CV was 12% for a concentration of 3.3 micrograms/L and 8.5% for 31.9 micrograms/L. The coefficient of correlation (r) with radioimmunoassay was 0.97. Curves for various dilutions of the macromolecular and monomeric forms of the hormone were not parallel.
Castracane, C. E., Cambiaso, C., Retegui, L. A., Gilbert, I., Ketelslegers, J.-M., & Masson, P. (1984). Particle-counting immunoassay of human somatotropin. Clinical Chemistry : international journal of molecular diagnostics and laboratory medicine, 30(5), 672-676. https://hdl.handle.net/2078.5/133298 (Original work published 1984)