Hémodialyse quotidienne à bas débit de dialysat à domicile : résultats cliniques et biologiques des 62 premiers patients traités en France et en Belgique

Benabed, Anaïs;Henri, Patrick;Lobbedez, Thierry;Goffin, Eric;Ficheux, Maxence;et.al.
(2017) Néphrologie & Thérapeutique — Vol. 13, n° 1, p. 18-25 (2017)

Files

BenabedetalHémodialysequotidienneàbasdébitdedialysatadomicilerésultatscliniquesetbiologiquesdes62premierspatientstraitésenFranceetenBelgiqueNephrolTher2017.pdf
  • Restricted Access
  • Adobe PDF
  • 660.61 KB

Details

Authors
  • Benabed, Anaïs
    Author
  • Henri, Patrick
    Author
  • Lobbedez, Thierry
    Author
  • Goffin, EricUCLouvain
    Author
  • Ficheux, Maxence
    Author
Show more
Abstract
(en) [Low flux dialysate daily home hemodialysis: A result for the 62 first French and Belgian patients] Since 2011, a new device is available for low flux dialysate quotidian home hemodialysis in France and Belgium. This study aims to evaluate the characteristics and dialysis prescriptions for Nx Stage System One™ users. We retrospectively included patients trained between 2011 and 2013 in France and Belgium. We collected data concerning their clinical features, their dialysis prescriptions, their laboratory parameters until 6 months of dialysis and, reason for dropping in case of cessation. Sixty-two patients from 31 centers, aged 48±18 years old, with a sex ratio 46/16 (M/F) are included with a median Charlson comorbidity index of 1 [0-3]. Of these patients, 71% are anuric and have been on dialysis for a mean time of 136.6±125 months. Previously, most of them had been taken care of in satellite units of dialysis (45%) and 14% are incident patients. In total, A total of 60% have an arterio-veinous fistula (AVF), with 18 patients using the Buttonhole system and 2 patients have a tunneled catheter. Median time for training was 26.5 days (17-45). Among the patients, 69% are dialyzed 6 days a week, during a mean time of 142.5±20 minutes with a volume of 20.9±3 liters of dialysate and without anticoagulant (63%). Predialytic levels of hemoglobin, creatinin, urea, phosphorus and β2microglobulin remain stable. On the contrary, there is a significant improvement of albumin and bicarbonate levels. Technique survival was 75% at 1 year, and major reason for cessation was kidney transplant. It seems that this device fits for young patients, with few comorbidities and a long past in renal chronic failure. These results suggest that dialysis adequacy is acceptable despite low dialysate volumes but need confirmation with a longer follow up and a larger cohort.
Affiliations

Citations

Benabed, A., Henri, P., Lobbedez, T., Goffin, E., Baluta, S., Benziane, A., Rachi, A., van der Pijl, J. W., Bechade, C., & Ficheux, M. (2017). Hémodialyse quotidienne à bas débit de dialysat à domicile : résultats cliniques et biologiques des 62 premiers patients traités en France et en Belgique. Néphrologie & Thérapeutique, 13(1), 18-25. https://doi.org/10.1016/j.nephro.2016.06.007 (Original work published 2017)