Fallacies and Pitfalls of Dialysis Sodium Prescription and Control

Stragier, André;Lopot, František;Švára, František;Polakovič, Vladimír
(2018) Blood Purification — Vol. 46, n° 1, p. 27-33 (2018)

Files

No attached file found for this publication.

Details

Authors
  • Stragier, AndréUCLouvain
    Author
  • Lopot, František
    Author
  • Švára, František
    Author
  • Polakovič, Vladimír
    Author
Abstract
Dialysate conductivity is routinely used as a surrogate for dialysate sodium concentration. However, dialysis machine manufacturers apply different conductivity temperature correction coefficients. With the same conductivity in dialysis machines manufactured by different manufacturers, dialysate sodium may significantly differ. Also, electrolyte prescriptions are individualized (K, Ca, HCO3) and this is associated with another variation in dialysate sodium in the order of 1–5 mmol/L and both deviations are cumulative and chronic for each patient. Equivalence of the prescribed dialysate sodium and the concentration measured in it is not granted. Both variables differ and it is machine dependent! This paper analyses those variations from a technical point of view and suggests how to detect them and how to deal with or avoid them in clinical practice.
Affiliations

Citations

Stragier, A., Lopot, F., Švára, F., & Polakovič, V. (2018). Fallacies and Pitfalls of Dialysis Sodium Prescription and Control. Blood Purification, 46(1), 27-33. https://doi.org/10.1159/000484921 (Original work published 2018)