Accurate distance measurement by an autonomous ultrasonic system combining time-of-flight and phase-shift methods

Gueuning, FE;Varlan, M;Eugene, CE;Dupuis, Pascal
(1997) IEEE Transactions on Instrumentation and Measurement — Vol. 46, n° 6, p. 1236-1240 (1997)

Files

No attached file found for this publication.

Details

Authors
  • Gueuning, FE
    Author
  • Varlan, M
    Author
  • Eugene, CE
    Author
  • Dupuis, PascalUCLouvain
    Author
Abstract
This paper presents an efficient algorithm for distance measurement, combining both the pulse time-of-flight method and the CW phase-shift method. It copes with a low-rate sampling technique allowed by the limited bandwidth of two ultrasonic transducers working in air at 30 kHz and with modest software resources of autonomous devices. The measuring system was implemented and tested on a compact Motorola MC68HC16-based platform, with a minimum of attached hardware. Experimental results show an accuracy better than 1 mm for a poor reflecting target at a distance of about 1 m.
Affiliations

Citations

Gueuning, F., Varlan, M., Eugene, C., & Dupuis, P. (1997). Accurate distance measurement by an autonomous ultrasonic system combining time-of-flight and phase-shift methods. IEEE Transactions on Instrumentation and Measurement, 46(6), 1236-1240. https://doi.org/10.1109/19.668260 (Original work published 1997)